If the photograph is clicked, it grows.

Fluke Y5020 20A 0.01ohm Calibration Shunt Resistance
SOLD OUT SOLD
par636






Nominal Resistance:0.01ohm+-1%,
Accuracy of Calibration Value:+-0.1%,
DC:+-0.1%,
DC to 1kHz:+-0.25%,
1kHz to 5kHz:+-0.35%,
Stability of Calibration Value:30ppm per 6 month,
Maximum Current:20A DC or RMS AC,
Burden Voltage at 20 AMPS:Less than 0.25V,
Temperature Coefficient:<=0.001% per degC,
power Coefficient:<=0.005% at 20AMPS,
Power:115VAC+-10% or 230VAC+-10%,11W for cooling fan,
Protection Class3