If the photograph is clicked, it grows.

Solartron 126096W 1260A 1296 Dielectric Measurement System
SOLD OUT SOLD
lcr207












1260A;
Generator;
Frequency Range:10uHz to 32MHz,
Error:+-100ppm, Stability:+-10ppm,
Amplitude;
<10MHz:0 to 3V,0 to 60mA,
>10Mhz:0 to 1V,0 to 20mA,
Resolution:5mV,100uA, Distortion:<2%,
DC Bias:+-40.95V,+-100mA
Sweep;
Types:Frequency,Amplitude,DC Bias,
Control:Up,Down,Step,Hold,
Maximum Current:+-100mA,
Maximum Voltage,Hi to Lo:+-46V,Lo to Ground:+-0.4V,
Connection:floting,single BNC,
Output is short-proof,

ANALYSIS;
Voltage Measurement;
Range/Resolution:30mV/1uV,300mV/10uV,3V/100uV,
Input Protected to:+-46V,
Input Configuration;
Connection;
Differential,BNC outers floating,
Differential,BNC outers grounded,
Single-ended,BNC outers floating,
Single-ended,BNC outers grounded,
Coupling:DC or AC(-3dB at 1Hz),
Impedance,Hi to outer:1Mohm+-2%,<35pF,
Common mode rejection(at 1MHz):>50dB,
Cross-channel isolation(at 1MHz):>100dB,
Noise Floor(at 1MHz):-110dBv,
Current Measurement;
Range/Resolution;
6uA/200pA,60uA/2nA,600uA/20nA,6mA/200nA,60mA/2uA,
Input Protected to:+-250mA,
Connection:floating,Single BNC,
Coupling:DC or AC(-3dB at 1Hz),
Impedance/Outer to Ground:100kohm/<200pF,
Outer floating to:+-0.4,
Integration time:10ms to 10E5 secs or auto,
Measurement delay:0 to 10E5secs,

Data Processing;
Math Operators:+,-,*,/,jw,powers,nested brackets,
Operand:V1,V2,I,Complex constans,
Statistical analysis:Standard diviation,variance,count,max,min,mean,

InterFace;
RS-232 and RS423,IEEE488(GP-IB),

Power require:90-126V,198-252V(selectable)@50/60Hz,
Dimensions:176mm(H)-432mm(W)-573mm(D),Weight:18kg,

12961A Dielectric constant measurement interface;
molecular liquid crystal dielectric relaxation process, such as high molecular
charge transport phenomena semiconductor, organic material, of ceramics
electrolytic capacitor, characterization of the electric double layer capacitor
Electrical, optical phenomenon of non-linear analysis
phenomenon analysis of electrode interface
phase transition, material structural analysis of the crystallization process
characterization of gas-liquid sensor

Impedance mesurement Range:100ohm to 100Tohm,
Frequency Range:10uHz to 10MHz,
Capaciatance Measurement Range:1pF to 0.1pF,
Tan Theta measurement Range:10-4 to 103,
Current Measurement Range:1fA to 100mA,
Bias dependence measurement of capacity(C-V),
The impedance measurement by AC voltage amplitude sweep
measurement of the complex relative permittivity

1260A+1296;
Frequency Range:10uHz to 10MHz,
Maximum Amplitude:7Vrms,
Naximum DC Bias:+-40V


Manual*7,Power Code,GP-IB Cable(408JE),BNC-BNC Cable(1m)*5,BNC-BNC Cable(50cm)*4,
Dielectric Reference Module(12961),
GP-IB-USB-HS,For ZPLOT USB SOFT KEY,CD-ROM(Scribnet Assocates,Advanced Software for Electrochemical Reserch and Development, Smart Impedance Software for 1269/1294/1296, ),
CD-ROM(NI-488.2 for Windows Vista/XP/2000),